High-performance, non-invasive magnetic stimulators for use in both the clinic and in medical research


MagPro R30 is a magnetic stimulator advanced high-performance, equipped with module "Option" for the delivery also of double stimulus in single mode or in a train of pairs. Its use is ideal in advanced diagnostics and research that use Transcranial Magnetic Stimulation (TMS) and Repetitive Transcranial Magnetic Stimulation (rTMS).


    • Examination of the physiology of the motor pathways in the central and peripheral nervous system

  • Is often used within Transcranial Magnetic Stimulation ( TMS ) and repetitive Transcranial Magnetic Stimulation ( rTMS ) research


  • Biphasic waveform
  • Stimulation rates up to 30 pulses per second
  • System operation control via a built-in computer, eliminating the need for an external computer to set up and control the timing of stimulus sequences
  • Flexible protocol storage in built-in computer
  • The MagPro R30 can be connected to an EMG system

The MagOption added to the MagPro R30 also gives following enhanced stimulation capabilities:

  • Dual pulses with up to 5 pulse pairs per second repetition rate
  • Biphasic and Monophasic
  • Dual up to 5pps both Biphasic and Monophasic


Performance data

Output versus Repetition Rate, depending on Waveform:



  mode wave form Repetition Rate
5 Hz 10 Hz 20 Hz 30 Hz



100 % 100 % 80 % 65 %
  standard monophasic 100 % 70 % 50 % 40 %
1 Hz 2 Hz 5 Hz
Twin and Dual Pulse Waveform


biphasic 100 % 100 % 100 %
Twin and Dual Pulse Waveform monophasic 100 % 100 % 70 %


Click here to see trigger details   trigger  

Trigger Input

  • Pulse width > 5μs
  • TTL + CMOS levels accepted
  • Input Impedance > 10 kOhm
  • Polarity: User-defined

Trigger Output

  • Pulse width: 30μs, TTL-levels
  • Output Impedance < 200 Ohm
  • Polarity: User-defined


Coils available

magpro coils

The magnetic stimulators line MagPro of Magventure have as many as 26 different types of coils to cover all applications diagnostic, therapeutic and research.


  • circular
  • butterfly
  • parabolic
  • elliptic
  • placebo
  • for MRI


The system based on liquid cooling allows sessions of repetitive stimulation in long duration without overheating problems