logo magventure nuovo lr trasp

magpro X100 option

MagPro X100 Option

 

double stimulation up to 100 Hz biphasic, monophasic and half sine for clinical and research top-performing

Description

MagPro X100 magnetic stimulator is an advanced, high-performance with module "Option" to deliver even in single or double stimulus trains in pairs, with biphasic, monophasic or sine. Its use is ideal in advanced diagnostics and research in which makes use of Transcranial Magnetic Stimulation (TMS) and Repetitive Transcranial Magnetic Stimulation (rTMS).

Applications

  • Examination of the physiology of the motor pathways in the central and peripheral nervous system
  • Examination of functional aspects of motor nerve stimulation
  • Examination of human cortical physiology
  • Often used within Transcranial Magnetic Stimulation ( TMS ) and repetitive Transcranial Magnetic Stimulation ( rTMS ) research

  Features

  • 4 waveforms: Biphasic, Biphasic Burst, Half Sine and Monophasic
  • Selectable current direction
  • Stimulation rates up to 100 pulses per second
  • Programmable input/output triggers
  • System operation control via a built-in computer, eliminating the need for an external computer to set up and control the timing of stimulus sequences
  • Flexible protocol storage in built-in computer

 

Performance data

Output versus Repetition Rate, depending on Waveform:

performance

 

  mode Wave form Repetition Rate
5 Hz 10 Hz 20 Hz 30 Hz 50 Hz 75 Hz 100Hz

  standard

biphasic

100 % 100 % 80 % 65 % 50% 35 % 30 %

  standard

half sine

100 % 100 % 80 % 65 % 50 % 35 % 30 %
  standard monophasic 100 % 70 % 50 % 40 % 30 % 25 % 20 %
1 Hz 2 Hz 5 Hz 10 Hz 15 Hz 20 Hz
  twin pulse bifasica 100 % 100 % 100 % 75 % 60 % 50 %
  twin pulse
half sine  100 % 100 % 100 % 75 % 60 % 50 %
  twin pulse monophasic 100 % 100 % 70 % 45 % 35 % 30 %

 

Click here to see trigger details   trigger  

Trigger Input

  • Pulse width > 5μs
  • TTL + CMOS levels accepted
  • Input Impedance > 10 kOhm
  • Polarity: User-defined

Trigger Output

  • Pulse width: 30μs, TTL-levels
  • Output Impedance < 200 Ohm 
  • Polarity: User-defined

 

Coils available

magpro coils

The magnetic stimulators line MagPro of Magventure have as many as 26 different types of coils to cover all applications diagnostic, therapeutic and research.

 

  • circular
  • butterfly
  • parabolic
  • elliptic
  • placebo
  • for MRI

 

 

The system based on liquid cooling allows sessions of repetitive stimulation in long duration without overheating problems.

MCF-B65 smallClick here to see the MagPro Coils page, containing coils detail

 

Product Product Request information form  euro small 

 